Photoacoustic determination of blood vessel diameter.

Loading...
Thumbnail Image

Files

Access status: Embargo until 2233-03-22 , Primary 57562thijssen.pdf (207.26 KB)

Date

Authors

Kolkman, R.G.
Klaessens, J.H.G.M. John
Hondebrink, E.
Hopman, J.C.W. Jeroen
Mul, F.F. de
Steenbergen, W.
Thijssen, J.M. Han
Leeuwen, T.G. van

Journal Title

Journal ISSN

Volume Title

Publisher

Research Projects

Organizational Units

Journal Issue

Abstract

A double-ring sensor was applied in photoacoustic tomographic imaging of artificial blood vessels as well as blood vessels in a rabbit ear. The peak-to-peak time (tau(pp)) of the laser (1064 nm) induced pressure transient was used to estimate the axial vessel diameter. Comparison with the actual vessel diameter showed that the diameter could be approximated by 2ctau(pp), with c the speed of sound in blood. Using this relation, the lateral diameter could also precisely be determined. In vivo imaging and monitoring of changes in vessel diameters was feasible. Finally, acoustic time traces were recorded while flushing a vessel in the rabbit ear with saline, which proved that the main contribution to the laser-induced pressure transient is caused by blood inside the vessel and that the vessel wall gives only a minor contribution.